Key product |
SMU Specifications Range 6 V Resolution and noise (0.1 Hz to 10 Hz) 6 μVExample of Calculating SMU Resolution The PXIe-4145 has a resolution of 100 ppm when set to a 100 μs aperture time. In the 6 V range, resolution can be calculated by multiplying 6 V by 100 ppm, as shown in the following equation: 6 V * 100 ppm = 6 V * 100 * 1×10-6 = 600 μV Likewise, in the 10 mA range, resolution can be calculated by multiplying 10 mA by 100 ppm, as shown in the following equation: 10 mA * 100 ppm = 10 mA * 100 * 1×10-6 = 1 μA |