Key product
Conditions
All specifications are valid under the following conditions unless otherwise noted.30 minutes warm-up time
Self-calibration is performed after the specified warm-up period has completed
Environment temperature is within the ambient range, and temperatures for individual PXIe-5820, PXIe-3622, PXIe-5653, and mmRH-5582 modules, as reported by their onboard temperature sensors, are within ±5 °C of the last self-calibration temperature, and temperature correction is enabled (default driver behavior)
Calibration cycle is maintained
Proper connector care and maintenance has been performed
Modules are installed in an NI chassis with slot cooling capacity = 82 W
The chassis fan mode is set to Auto and Cooling Profile is set to 58 W/82 W in NI Measurement & Automation Explorer (MAX)
Empty chassis slots contain slot blockers and EMC filler panels to minimize temperature drift and reduce emissions
Modules are connected with NI cables as shown in the PXIe-5831 Getting Started Guide
RFmx, NI-RFSA 19.6 or later, or NI-RFSG 19.6 or later instrument driver is used, and driver default settings are used unless otherwise noted
Calibration IP is used properly during the creation of custom FPGA bitfiles
LO Step Size is set to the default value and the LO Source is set to Onboard
Acquisition Type is set to IQ
Warranted specifications are valid under the following condition unless otherwise noted.Over ambient temperature ranges of 0 °C to 45 °C for IF ports
Over ambient temperature ranges of 23 °C± 5 °C for RF ports
Typical and Typical-95 specifications are valid under the following condition unless otherwise noted.Over ambient temperature ranges of 23 °C± 5 °C for IF and RF ports